Environmental Monitoring Procedure Using the Radiation Alert® GammaView

Environmental Monitoring Procedure Using the Radiation Alert® GammaView

The Radiation Alert® GammaView is purpose-built for sensitive gamma radiation detection. Its NaI scintillation detector provides the responsiveness needed to identify subtle changes in low-level gamma fields, making the instrument particularly well suited for environmental surveillance, background radiation trending, site boundary monitoring, NORM assessments, and post-incident environmental surveys.

The GammaView is specifically marketed for environmental monitoring, compliance monitoring, health physics, and remote monitoring applications.

  1. Purpose

The purpose of this procedure is to establish a standardized method for conducting environmental gamma radiation monitoring using the Radiation Alert® GammaView to:

  • Establish environmental baseline radiation levels.
  • Detect changes in gamma radiation fields.
  • Evaluate radiological impacts from licensed activities.
  • Monitor facility boundaries.
  • Support regulatory compliance programs.
  • Investigate abnormal radiation readings.
  • Trend long-term environmental conditions.
  1. Capabilities of the GammaView

The GammaView utilizes a 1″ × 1″ or 2”x 2” NaI scintillation detector, providing substantially greater sensitivity to gamma radiation than a typical GM survey meter. It measures gamma dose rates, count rates, and total counts and is capable of detecting small changes in environmental background radiation.

Typical Operating Ranges

  • Dose rate: 0.001 µSv/hr to 50 µSv/hr
  • Exposure rate: 1 µR/hr to 5 mR/hr

Approximate Gamma Energy Range: 20 keV to 1.5 MeV, depending on probe configuration.

  1. Safety Requirements

Personnel shall:

  • Complete radiation safety training.
  • Follow facility ALARA principles.
  • Wear required PPE when entering controlled areas.
  • Avoid unnecessary exposure.
  • Follow site-specific emergency procedures.

The GammaView is a monitoring instrument and should not be relied upon as a personal dosimeter.

  1. Instrument Verification

Prior to use:

Visual Inspection

Verify:

  • Detector housing intact
  • Cable undamaged
  • Connectors secure
  • Display functioning properly

Calibration

Confirm calibration is current.

Battery Check

Verify battery levels are acceptable.

Functional Check

Measure a known background location and verify instrument response is normal for the area.

The manufacturer’s operation manual recommends connecting or disconnecting the detector only when the instrument is powered off.

  1. Establishing Environmental Baseline

Before beginning a monitoring program:

  1. Select permanent monitoring locations.
  2. Obtain readings under normal operating conditions.
  3. Measure each location multiple times.
  4. Calculate average background values.

Record:

  • Location ID
  • GPS coordinates
  • Date
  • Time
  • Weather conditions
  • Dose rate (µSv/hr)
  • Count rate (CPM)
  • Technician

Baseline values become the comparison standard for future surveys.

  1. Fixed Environmental Monitoring Locations

Typical locations include:

Site Boundary Stations

  • North boundary
  • South boundary
  • East boundary
  • West boundary

Public Access Areas

  • Parking lots
  • Visitor centers
  • Adjacent properties

Sensitive Environmental Areas

  • Surface water discharge points
  • Stormwater drainage channels
  • Wetlands
  • Wildlife areas

Facility Areas

  • Waste storage locations
  • Source storage facilities
  • Material handling areas
  1. Static Environmental Survey Procedure

This is the preferred method for trend analysis.

Setup

Position the detector:

  • 1 meter above the ground
  • Away from vehicles
  • Away from buildings, when possible

Measurement

Allow the detector to stabilize.

Collect readings for:

Recommended collection times:

  • Minimum: 1 minute
  • Preferred: 5 minutes
  • High-sensitivity surveys: 10 minutes

Record the following:

  • Average CPM
  • Maximum CPM
  • Dose rate
  • Time
  • Weather conditions

Because the GammaView is highly sensitive, using longer count times significantly improves environmental data quality.

  1. Environmental Walk-Over Surveys

Walk-over surveys are useful for locating unexpected radioactive material.

Survey Method

Detector Height

Hold the detector 0.5 to 1 meter above the ground.

Walking Speed

Walk slowly at approximately 1–2 mph.

Watch for:

  • Count-rate increases
  • Elevated dose rates
  • Audible alarm activation

Investigate any readings significantly above established background levels.

  1. Soil Monitoring

For environmental screening:

Grid Survey Method

Establish a grid.

Example:

  • 25-meter spacing
  • 50-meter spacing
  • 100-meter spacing

At each point:

  1. Measure at 1 meter above grade.
  2. Measure at ground level.
  3. Record CPM and dose rate.

Any statistically significant elevation above baseline should be investigated further.

  1. Surface Water Monitoring

The GammaView cannot directly quantify radionuclides in water, but it can identify elevated gamma activity associated with:

  • Sediments
  • Filter residues
  • Shoreline deposits

Procedure:

  1. Collect sediment or filter samples.
  2. Place in a repeatable counting geometry.
  3. Count samples for a fixed interval.
  4. Compare against control samples.
  1. Vegetation Monitoring

Environmental surveillance programs often use:

  • Grass
  • Leaves
  • Moss
  • Agricultural products

Samples should:

  1. Be collected consistently.
  2. Be counted using the same geometry.
  3. Be measured for a fixed counting interval.

The GammaView’s scintillation detector is extremely effective at identifying subtle gamma activity from deposited radionuclides.

  1. Post-Incident Environmental Monitoring

Following a release, spill, or contamination event:

Initial Survey

Establish:

  • Impacted area
  • Radiation gradient
  • Potential migration pathways

Follow-Up Monitoring

Survey:

  • Drainage pathways
  • Soil
  • Vegetation
  • Building surfaces
  • Public boundaries

Map all measurements to identify contamination spread.

  1. Use of the Single Channel Analyzer (SCA)

One unique capability of the GammaView is its Single Channel Analyzer (SCA) mode.

This allows users to create an energy “window” that focuses on specific gamma energies while reducing background response. This can improve sensitivity to particular radionuclides in the field and reduce interference from naturally occurring background radiation.

Potential uses include:

  • Environmental investigations
  • Source searches
  • NORM discrimination
  • Focused surveillance of known radionuclides
  1. Investigation Levels

Example environmental action levels:

Condition Response
Within baseline range Record only
2× normal background Repeat measurement
3× normal background Notify supervisor
Persistent elevated readings RSO investigation
Confirmed abnormal radiation field Implement corrective actions

Actual investigation levels should be established by the Radiation Safety Officer.

  1. Data Trending

Environmental monitoring programs should develop long-term trend analyses.

Track:

  • Monthly averages
  • Quarterly averages
  • Seasonal variations
  • Rainfall impacts
  • Operational changes
  • Construction activities

Because the GammaView is highly sensitive to low-level gamma fields, it is well suited for detecting long-term changes that may not be noticeable with standard GM-based survey meters.

  1. Limitations

The GammaView:

Can Do

  • Detect very low gamma radiation levels.
  • Establish environmental baselines.
  • Trend environmental conditions.
  • Monitor facility boundaries.
  • Perform compliance monitoring.

Cannot Do

  • Positively identify radionuclides by itself.
  • Replace laboratory gamma spectroscopy.
  • Measure alpha contamination directly.
  • Be used as a personal dosimeter.

Recommended Use

For a formal environmental radiation monitoring program, the GammaView is generally better suited than the Ranger. While the Ranger excels in contamination surveys and wipe testing, the GammaView’s NaI scintillation detector offers the sensitivity required for environmental background monitoring, perimeter surveillance, NORM investigations, and post-incident radiological assessments.

 